VB LabSolutions

SORRY, THIS PAGE IS UNDER CONSTRUCTION....

Low Response or Decreasing Response

Note: We observed several times 90% problem was related to sample introduction area, turbo ion source side.



  1. Decreasing response observe. 
  2. Low response, or No response observe.
  3. Uneven response observe.
 Technical Observation:
  1. Turbo ion V source electrode or peak union was found partially or fully blocked.
  2. Microscopic inspection so many time found that the electrode tip was brunt or block.
  3. Same symptoms observed, but root cause was found different. Hence it's very difficult to diagnose. And there for needs to taken some precaution.

Check List 
  1. Electrode.
  2. Peak union.
  3. SS splitter or union.
  4. Peak tubing.
  5. Probe Nozzle blockage or damage.
  6. Peek ferrule leakage.
  7. Electrode leakage inside to probe.
  8. Orifice plate hole blockage by lent from tissue paper during cleaning orifice plate.
  9. Contamination and Buffer deposition on curtain and Orifice plate.
  10. Gas leakage or blockage exhaust line clogged.
  11. Loose ferrule courses mobile phase leakage and high voltage arcing turbo ion source damage.

Thank You For Your Attentions..
For More Tips Get In Teach With Us...

IF YOUR NOT TAKING SERIOUSLY, THINK AGAIN OR READY TO PAY FOR IT

Note: We observed several times 90% problem was related to sample introduction area, turbo ion source side.



  1. Decreasing response observe. 
  2. Low response, or No response observe.
  3. Uneven response observe.
 Technical Observation:
  1. Turbo ion V source electrode or peak union was found partially or fully blocked.
  2. Microscopic inspection so many time found that the electrode tip was brunt or block.
  3. Same symptoms observed, but root cause was found different. Hence it's very difficult to diagnose. And there for needs to taken some precaution.

Check List 
  1. Electrode.
  2. Peak union.
  3. SS splitter or union.
  4. Peak tubing.
  5. Probe Nozzle blockage or damage.
  6. Peek ferrule leakage.
  7. Electrode leakage inside to probe.
  8. Orifice plate hole blockage by lent from tissue paper during cleaning orifice plate.
  9. Contamination and Buffer deposition on curtain and Orifice plate.
  10. Gas leakage or blockage exhaust line clogged.
  11. Loose ferrule courses mobile phase leakage and high voltage arcing turbo ion source damage.

Thank You For Your Attentions..
For More Tips Get In Teach With Us...

Buffer Deposition 

Note: We observed several times 90% problem was related to sample introduction area, turbo ion source side.



  1. Decreasing response observe. 
  2. Low response, or No response observe.
  3. Uneven response observe.
 Technical Observation:
  1. Turbo ion V source electrode or peak union was found partially or fully blocked.
  2. Microscopic inspection so many time found that the electrode tip was brunt or block.
  3. Same symptoms observed, but root cause was found different. Hence it's very difficult to diagnose. And there for needs to taken some precaution.

Check List 
  1. Electrode.
  2. Peak union.
  3. SS splitter or union.
  4. Peak tubing.
  5. Probe Nozzle blockage or damage.
  6. Peek ferrule leakage.
  7. Electrode leakage inside to probe.
  8. Orifice plate hole blockage by lent from tissue paper during cleaning orifice plate.
  9. Contamination and Buffer deposition on curtain and Orifice plate.
  10. Gas leakage or blockage exhaust line clogged.
  11. Loose ferrule courses mobile phase leakage and high voltage arcing turbo ion source damage.

Thank You For Your Attentions..
For More Tips Get In Teach With Us...

GAS LEAKAGE

Note: We observed several times 90% problem was related to sample introduction area, turbo ion source side.



  1. Decreasing response observe. 
  2. Low response, or No response observe.
  3. Uneven response observe.
 Technical Observation:
  1. Turbo ion V source electrode or peak union was found partially or fully blocked.
  2. Microscopic inspection so many time found that the electrode tip was brunt or block.
  3. Same symptoms observed, but root cause was found different. Hence it's very difficult to diagnose. And there for needs to taken some precaution.

Check List 
  1. Electrode.
  2. Peak union.
  3. SS splitter or union.
  4. Peak tubing.
  5. Probe Nozzle blockage or damage.
  6. Peek ferrule leakage.
  7. Electrode leakage inside to probe.
  8. Orifice plate hole blockage by lent from tissue paper during cleaning orifice plate.
  9. Contamination and Buffer deposition on curtain and Orifice plate.
  10. Gas leakage or blockage exhaust line clogged.
  11. Loose ferrule courses mobile phase leakage and high voltage arcing turbo ion source damage.

Thank You For Your Attentions..
For More Tips Get In Teach With Us...

Mobile Phase Leakage & High Volt Arcing

Note: We observed several times 90% problem was related to sample introduction area, turbo ion source side.



  1. Decreasing response observe. 
  2. Low response, or No response observe.
  3. Uneven response observe.
 Technical Observation:
  1. Turbo ion V source electrode or peak union was found partially or fully blocked.
  2. Microscopic inspection so many time found that the electrode tip was brunt or block.
  3. Same symptoms observed, but root cause was found different. Hence it's very difficult to diagnose. And there for needs to taken some precaution.

Check List 
  1. Electrode.
  2. Peak union.
  3. SS splitter or union.
  4. Peak tubing.
  5. Probe Nozzle blockage or damage.
  6. Peek ferrule leakage.
  7. Electrode leakage inside to probe.
  8. Orifice plate hole blockage by lent from tissue paper during cleaning orifice plate.
  9. Contamination and Buffer deposition on curtain and Orifice plate.
  10. Gas leakage or blockage exhaust line clogged.
  11. Loose ferrule courses mobile phase leakage and high voltage arcing turbo ion source damage.

Thank You For Your Attentions..
For More Tips Get In Teach With Us...

Exhaust  Line Clogged

Note: We observed several times 90% problem was related to sample introduction area, turbo ion source side.



  1. Decreasing response observe. 
  2. Low response, or No response observe.
  3. Uneven response observe.
 Technical Observation:
  1. Turbo ion V source electrode or peak union was found partially or fully blocked.
  2. Microscopic inspection so many time found that the electrode tip was brunt or block.
  3. Same symptoms observed, but root cause was found different. Hence it's very difficult to diagnose. And there for needs to taken some precaution.

Check List 
  1. Electrode.
  2. Peak union.
  3. SS splitter or union.
  4. Peak tubing.
  5. Probe Nozzle blockage or damage.
  6. Peek ferrule leakage.
  7. Electrode leakage inside to probe.
  8. Orifice plate hole blockage by lent from tissue paper during cleaning orifice plate.
  9. Contamination and Buffer deposition on curtain and Orifice plate.
  10. Gas leakage or blockage exhaust line clogged.
  11. Loose ferrule courses mobile phase leakage and high voltage arcing turbo ion source damage.

Thank You For Your Attentions..
For More Tips Get In Teach With Us...

Orifice Plate Hole Blocked By Lent 

Note: We observed several times 90% problem was related to sample introduction area, turbo ion source side.



  1. Decreasing response observe. 
  2. Low response, or No response observe.
  3. Uneven response observe.
 Technical Observation:
  1. Turbo ion V source electrode or peak union was found partially or fully blocked.
  2. Microscopic inspection so many time found that the electrode tip was brunt or block.
  3. Same symptoms observed, but root cause was found different. Hence it's very difficult to diagnose. And there for needs to taken some precaution.

Check List 
  1. Electrode.
  2. Peak union.
  3. SS splitter or union.
  4. Peak tubing.
  5. Probe Nozzle blockage or damage.
  6. Peek ferrule leakage.
  7. Electrode leakage inside to probe.
  8. Orifice plate hole blockage by lent from tissue paper during cleaning orifice plate.
  9. Contamination and Buffer deposition on curtain and Orifice plate.
  10. Gas leakage or blockage exhaust line clogged.
  11. Loose ferrule courses mobile phase leakage and high voltage arcing turbo ion source damage.

Thank You For Your Attentions..
For More Tips Get In Teach With Us...

Orifice Plate Hole Blocked By Lent 

Note: We observed several times 90% problem was related to sample introduction area, turbo ion source side.



  1. Decreasing response observe. 
  2. Low response, or No response observe.
  3. Uneven response observe.
 Technical Observation:
  1. Turbo ion V source electrode or peak union was found partially or fully blocked.
  2. Microscopic inspection so many time found that the electrode tip was brunt or block.
  3. Same symptoms observed, but root cause was found different. Hence it's very difficult to diagnose. And there for needs to taken some precaution.

Check List 
  1. Electrode.
  2. Peak union.
  3. SS splitter or union.
  4. Peak tubing.
  5. Probe Nozzle blockage or damage.
  6. Peek ferrule leakage.
  7. Electrode leakage inside to probe.
  8. Orifice plate hole blockage by lent from tissue paper during cleaning orifice plate.
  9. Contamination and Buffer deposition on curtain and Orifice plate.
  10. Gas leakage or blockage exhaust line clogged.
  11. Loose ferrule courses mobile phase leakage and high voltage arcing turbo ion source damage.

Thank You For Your Attentions..
For More Tips Get In Teach With Us...

Orifice Plate Hole Blocked By Lent 

Note: We observed several times 90% problem was related to sample introduction area, turbo ion source side.



  1. Decreasing response observe. 
  2. Low response, or No response observe.
  3. Uneven response observe.
 Technical Observation:
  1. Turbo ion V source electrode or peak union was found partially or fully blocked.
  2. Microscopic inspection so many time found that the electrode tip was brunt or block.
  3. Same symptoms observed, but root cause was found different. Hence it's very difficult to diagnose. And there for needs to taken some precaution.

Check List 
  1. Electrode.
  2. Peak union.
  3. SS splitter or union.
  4. Peak tubing.
  5. Probe Nozzle blockage or damage.
  6. Peek ferrule leakage.
  7. Electrode leakage inside to probe.
  8. Orifice plate hole blockage by lent from tissue paper during cleaning orifice plate.
  9. Contamination and Buffer deposition on curtain and Orifice plate.
  10. Gas leakage or blockage exhaust line clogged.
  11. Loose ferrule courses mobile phase leakage and high voltage arcing turbo ion source damage.

Thank You For Your Attentions..
For More Tips Get In Teach With Us...